• Detection mode: Fluorescence, Luminescence, UV-Vis absorbance and Time-resolved fluorescence
• Light sources: Xenon flash
• Detector: PMT for florescence system, Photodiode for absorbance
• Wavelength range: 250-700nm (fluorescence), 230-999nm (absorbance)
• Dynamic range: 7 decades (fluorescence), 0.4 OD (absorbance), >6 decades (luminescence)
• Reading speed (kinetic): 96 wells: 11 seconds, 384 wells: 22 seconds
• Assay type: End point, kinetic, spectral and well scanning read
• Plate compatibility: Reads 6 to 384 well plates and imaging of 6 to 1536 well plates. we can accommodate microscope slides, petri and cell culture dishes, cell culture flasks (T25)
• Plate shaking: Linear, orbital, double orbital shaking modes
• Imaging modes: Fluorescence, bright-field, color bright-field and phase contrast with single color, multi-color, montage, time lapse and z-stacking
• Image filter cube: DAPI, GFP, Texas Red Filter cubes with imaging LED cubes of 365nm, 465nm and 590nm
• Monochromator wavelength accuracy and repeatability: 2 nm, +0.2 nm, OD accuracy: <1% at 2.0 OD <3% at 3.0 OD, OD linearity; <1% from 0 to 3.0 OD, OD repeatability: <0.5% at 2.0 OD